FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 3823034 · Received February 5, 2014

Report

Report Number
1720753-2014-01172
Event Type
Malfunction
Date Received
February 5, 2014
Date of Event
December 9, 2013
Report Date
February 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE CPU BATTERY NEEDS TO BE REPLACED. THE REPORTED ISSUE IS CURRENTLY UNDER INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MDA MEMORY PREVENTED THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
75949 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1