FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 3823007 · Received February 19, 2014

Report

Report Number
9615050-2014-01275
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
January 6, 2014
Report Date
January 7, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING AT THE SERVICE CTR WHEN POWERED ON THE DEVICE DISPLAYED A WHITESCREEN ERROR, "DEVICE MALFUNCTION: HW WATCHDOG ERROR AND UNEXPECTED WATCHDOG RESET." THIS WAS DUE TO A DEPLETED BACK UP BATTERY ON THE USER INTERFACE CONTROLLER (UIC). THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING VERIFICATION TESTING AT THE SERVICE CTR, AFTER THE DEVICE WAS POWERED ON, THE DEVICE ALARMED WITH A WHITESCREEN ERROR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104265 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA