FDA Adverse Event
Malfunction
Summary report: N
LC PCA NEW 802.11 AB
MDR report key: 3822897
·
Received February 5, 2014
Report
- Report Number
- 9615050-2014-00914
- Event Type
- Malfunction
- Date Received
- February 5, 2014
- Date of Event
- January 9, 2014
- Report Date
- January 9, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- MEA
- PMA / PMN Number
- K042800
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THIS WAS DUE TO A PIN THAT WAS LIFTED FROM THE CONDUCTOR TRACE ON THE BOTTOM OF THE AUDIO TRANSDUCER WHICH DISABLED THE OUTPUT. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE INTERMITTENTLY DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76436 | LC PCA NEW 802.11 AB | 80MEA | MEA | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |