FDA Adverse Event Malfunction Summary report: N

SYMBIQ DCHANNEL 3 01

MDR report key: 3822878 · Received February 5, 2014

Report

Report Number
9615050-2014-00926
Event Type
Malfunction
Date Received
February 5, 2014
Date of Event
January 1, 2014
Report Date
January 6, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE PASSED TESTING. DURING A REVIEW OF THE DEVICE HISTORY AT THE SERVICE CENTER, A S321 (MOTOR ERROR-PMC, LEFT) WAS NOTED. ALTHOUGH THE DEVICE PASSED TESTING, THE PROBABLE CAUSE OF THE CUSTOMER'S REPORTED S321 MALFUNCTION ALARM CODE WAS DUE TO A BROKEN ENCODER ON THE CHANNEL A PLUNGER MOTOR. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH A S321 (MOTOR ERROR - PMC, LEFT) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFO WAS PROVIDED; THEREFORE SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76357 SYMBIQ DCHANNEL 3 01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK