FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3822864 · Received February 5, 2014

Report

Report Number
8020893-2014-00273
Event Type
Malfunction
Date Received
February 5, 2014
Date of Event
January 1, 2014
Report Date
January 10, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

COVIDIEN REFERENCE NUMBER: (B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT WHEN AN 840 VENTILATOR WAS POWERED ON THE DISPLAY WOULD FLASH AND THEN GO BLANK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.