FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3822864
·
Received February 5, 2014
Report
- Report Number
- 8020893-2014-00273
- Event Type
- Malfunction
- Date Received
- February 5, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 10, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
COVIDIEN REFERENCE NUMBER: (B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Description of Event or Problem · 1
COVIDIEN RECEIVED INFORMATION STATING THAT WHEN AN 840 VENTILATOR WAS POWERED ON THE DISPLAY WOULD FLASH AND THEN GO BLANK. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76364 | 840 VENTILATOR | CBK: VENTILATOR, CONTINUOUS, FACILITY | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |