FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3822846 · Received February 5, 2014

Report

Report Number
8020893-2014-00283
Event Type
Malfunction
Date Received
February 5, 2014
Date of Event
January 13, 2014
Report Date
January 13, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHIC USER INTERFACE (GUI) LIQUID CRYSTAL DISPLAY (LCD). THE UNIT PASSED EXTENDED SELF-TESTING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD WHITE FUZZY BOTTOM SCREEN. THE FAILURE HAPPENED WHILE THE DEVICE WAS NOT USED ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76304 840 VENTILATOR CBK: VENTILATOR, CONTINUOUS, FACILITY CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1