FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUALCHAN INF

MDR report key: 3822781 · Received February 7, 2014

Report

Report Number
9615050-2014-00994
Event Type
Malfunction
Date Received
February 7, 2014
Date of Event
January 1, 2014
Report Date
January 8, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0070-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO RESPOND INTERMITTENTLY. THIS WAS DUE TO FLUID INGRESS FROM THE PRODUCT USED TO CLEAN THE DEVICE TOUCHSCREEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80856 SYMBIQ DUALCHAN INF 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA