FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ DUAL CHANNEL
MDR report key: 3822778
·
Received February 7, 2014
Report
- Report Number
- 9615050-2014-00987
- Event Type
- Malfunction
- Date Received
- February 7, 2014
- Date of Event
- November 1, 2013
- Report Date
- January 9, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND AT POWER UP, THE DEVICE DISPLAYED A WHITESCREEN ERROR WITHOUT AN AUDIBLE ALARM TONE. THE PROBABLE CAUSE WAS DUE TO A LOOSE RIBBON CABLE CONNECTED TO THE FRONT DISPLAY ON THE UIC (USER INTERFACE CONTROLLER). THE LOOSE CABLE DOES NOT TRIGGER AN ALARM CONDITION. THE CAUSE OF THE LOOSE CABLE COULD NOT BE DETERMINED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 80855 | SYMBIQ DUAL CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |