FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3822754 · Received February 7, 2014

Report

Report Number
1720753-2014-01267
Event Type
Malfunction
Date Received
February 7, 2014
Date of Event
January 28, 2014
Report Date
February 7, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE VOLTAGE ON PS1 POWER SUPPLY WAS FOUND TO BE LOW AND WAS ADJUSTED TO OPTIMUM OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TOT HIS EVENT.