SYNERGY?
Report
- Report Number
- 2134265-2014-02823
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- April 26, 2014
- Report Date
- April 28, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ES
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS COMBINATION PRODUCT. (B)(4).
DEVICE EVALUATED BY MANUFACTURER: A VISUAL EXAMINATION OF THE RETURNED DEVICE IDENTIFIED A KINK IN THE HYPOTUBE AT 12CM DISTAL TO THE STRAIN RELIEF WHICH IS CONSISTENT WITH EXCESSIVE FORCE APPLIED TO THE HYPOTUBE DURING HANDLING. AN EXAMINATION OF THE CRIMPED STENT IDENTIFIED THAT THE PROXIMAL STENT STRUT ROWS OF THE STENT WERE SECURELY CRIMPED ONTO THE BALLOON. HOWEVER THE MID TO DISTAL SECTION OF THE STENT WAS STRETCHED AND PULLED DISTALLY OVER THE TIP. THERE WAS NO EVIDENCE OF ANY DEVICE USE. NO INFLATION MEDIA WAS PRESENT INSIDE THE INFLATION LUMEN. THE STENT PROTECTOR WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT (B)(4).
IT WAS REPORTED THAT DURING PREPARATION OF A STENTING PROCEDURE, STENT DAMAGE OCCURRED. STENT DAMAGE WAS NOTED ON THE DISTAL TIP OF A 2.5X20MM SYNERGY WHILE IT WAS BEING REMOVED FROM THE CARRIER TUBE DURING PREPARATION. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PROCEDURE AS COMPLETED WITH ANOTHER SAME DEVICE. THE PATIENT'S STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED U.S. DEVICE.
IT WAS REPORTED THAT DURING PREPARATION OF A STENTING PROCEDURE, STENT DAMAGE OCCURRED. STENT DAMAGE WAS NOTED ON THE DISTAL TIP OF A 2.5X20MM SYNERGY WHILE IT WAS BEING REMOVED FROM THE CARRIER TUBE DURING PREPARATION. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PROCEDURE AS COMPLETED WITH ANOTHER SAME DEVICE. THE PATIENT'S STATUS IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303742 | SYNERGY? | CORONARY DRUG-ELUTING STENT | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493926220250 | 16437820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |