FDA Adverse Event Malfunction Summary report: N

ULTRAFLEX? TRACHEOBRONCHIAL

MDR report key: 3822491 · Received May 21, 2014

Report

Report Number
3005099803-2014-02033
Event Type
Malfunction
Date Received
May 21, 2014
Date of Event
April 4, 2014
Report Date
April 30, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
JCT
PMA / PMN Number
K121048
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION REPORT OF STENT PARTIALLY DEPLOYED. A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE STENT WAS PARTIALLY DEPLOYED BY 20MM. A BEND WAS NOTED IN THE SHAFT PROXIMAL TO STENT. DURING ANALYSIS IT WAS POSSIBLE TO RETRACT THE DEPLOYMENT SUTURE AND FULLY DEPLOY THE STENT WITHOUT ISSUE. NO ISSUES NOTED WITH STENT. THE INVESTIGATION CONCLUDED THAT THIS COMPLAINT IS ASSOCIATED WITH A PRODUCT THAT MEETS DESIGN AND MANUFACTURE SPECIFICATIONS BUT DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, PERFORMANCE OF THE DEVICE WAS LIMITED.  THEREFORE, THE MOST PROBABLE ROOT CAUSE CLASSIFICATION FOR THE REPORTED FAILURE IS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. A LABELING REVIEW WAS PERFORMED, AND FROM THE INFORMATION AVAILABLE THIS DEVICE WAS USED PER THE DIRECTIONS FOR USE (DFU) / PRODUCT LABEL.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX TRACHEOBRONCHIAL STENT WAS USED IN THE AIRWAY DURING A BRONCHOSCOPY WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, THE DEPLOYMENT SUTURE CANNOT BE RELEASED AND THE CATHETER BUCKLED. THE PHYSICIAN REMOVED THE DEVICE FULLY CONSTRAINED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX TRACHEOBRONCHIAL STENT WAS USED IN THE AIRWAY DURING A BRONCHOSCOPY WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, THE DEPLOYMENT SUTURE CANNOT BE RELEASED AND THE CATHETER BUCKLED. THE PHYSICIAN REMOVED THE DEVICE FULLY CONSTRAINED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR-REPORTABLE EVENT BASED ON INVESTIGATION RESULTS WHICH REVEALED THE STENT WAS RECEIVED PARTIALLY DEPLOYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303736 ULTRAFLEX? TRACHEOBRONCHIAL PROSTHESIS, TRACHEAL, EXPANDABLE JCT BOSTON SCIENTIFIC - GALWAY M00576300 16566630

Patients

Seq Age Sex Outcome Treatment
1