FDA Adverse Event Malfunction Summary report: N

MUSTANG?

MDR report key: 3822458 · Received May 20, 2014

Report

Report Number
2134265-2014-02708
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 18, 2014
Report Date
April 22, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
FGE
PMA / PMN Number
K110122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: A VISUAL AND TACTILE EXAMINATION OF THE RETURNED DEVICE FOUND NO KINKS OR DAMAGE ALONG THE ENTIRE LENGTH OF THE DEVICE. A FURTHER MICROSCOPIC ANALYSIS REVEALED LONGITUDINAL TEAR IN THE MID BALLOON BODY MEASURING APPROXIMATELY 50MM IN LENGTH. THE EXAMINATION OF THE BALLOON MATERIAL AND MARKERBANDS IDENTIFIED NO ISSUES WHICH COULD POTENTIALLY HAVE CONTRIBUTED TO THIS COMPLAINT. NO OTHER DEFECTS WERE NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED CEPHALIC VEIN. A 7.0MM X 40MM, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON THE FIRST INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED CEPHALIC VEIN. A 7.0MM X 40MM, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, ON THE FIRST INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED INTACT FROM THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300506 MUSTANG? CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - GALWAY H74939171070470 16629625

Patients

Seq Age Sex Outcome Treatment
1 INTRODUCER SHEATH: 5FR / TERUMO| INFLATION DEVICE: KANEKA| GUIDE WIRE: RADIFOCUS / TERUMO