FDA Adverse Event Malfunction Summary report: N

EMERGE?

MDR report key: 3822446 · Received May 20, 2014

Report

Report Number
2134265-2014-02673
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 22, 2014
Report Date
April 22, 2014
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
K113220
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR: THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED, CHRONIC TOTAL OCCLUSION (CTO) TARGET LESION WAS LOCATED IN A SEVERELY CALCIFIED AND MODERATELY TORTUOUS DISTAL LEFT CIRCUMFLEX (LCX). A 2.00MM X 15MM EMERGE¿ CATHETER BALLOON WAS ADVANCED FOR PRE-DILATATION AND THE DEVICE WAS INFLATED TWICE AT 14 ATMOSPHERES. HOWEVER, ON THE THIRD INFLATION AT 14 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH A 1.5MM X 15MM EMERGE¿ BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299638 EMERGE? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493919315200 16523694

Patients

Seq Age Sex Outcome Treatment