FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3822351 · Received May 20, 2014

Report

Report Number
2531779-2014-14252
Event Type
Malfunction
Date Received
May 20, 2014
Report Date
May 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 07/09/2014 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/01/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE PUMP POWERED UP TO THE VERIFY SCREEN WITH THE APPROPRIATE AUDIBLE AND VIBRATORY FEATURES. IT WAS OBSERVED THAT THE VERIFY SCREEN WAS DIM AND THE TEXT WAS DISCOLORED. THE REPORTED BLANK DISPLAY ISSUE WAS NOT DUPLICATE DURING THE INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. REPORTEDLY, THE PUMP POWERED UP TO A BLANK DISPLAY SCREEN. THE REPORTER STATED THAT THERE WAS NO EVIDENCE OF MOISTURE OR CORROSION IN THE PUMP AND CONTRAST RESET DID NOT RESOLVE THE ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
298356 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 20 YR