UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2122870-2014-00392
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 22, 2014
- Report Date
- April 28, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) OBSERVED A 10-TOOTH MIXER PULLEY WITH A FLATTENED RING. THE FSE REPLACED THE MIXER PULLEY, ASPIRATE PROBES AND PERISTALTIC PUMP TUBING. THE FSE REMOVED AND CLEANED THE WASH WHEEL AND INSPECTED THE BEARINGS. THE FSE COMPLETED PASSING SYSTEM CHECK AND HIGH SENSITIVITY (HS) SYSTEM CHECK AND PERFORMED A PRECISION TEST WHICH FAILED. THE FSE REPLACED THE PIPETTOR TIP AND MAJOR COMPONENTS ON THE ANALYTICAL MODULE AND PERFORMED ALL NECESSARY ALIGNMENTS. IN ADDITION, THE FSE ADJUSTED THE ULTRASONICS AND MIXER SPEED. THE FSE PERFORMED A PASSING SYSTEM CHECK, HS SYSTEM CHECK, AND PRECISION TEST. SERVICE ACTIVITY PERFORMED MET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. IN CONCLUSION, SYSTEM HARDWARE IS THE LIKELY CAUSE OF THE EVENT AS REPAIRS AND ADJUSTMENTS RESOLVED THE REPORTED ISSUE. HOWEVER, A SPECIFIC HARDWARE COMPONENT WAS NOT IDENTIFIED AS THE SINGULAR CAUSE OF THE EVENT. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS INCIDENT: 2122870-2014-00391, 2122870-2014-00392, 2122870-2014-00393, 2122870-2014-00394.
THE CUSTOMER REPORTED FALSELY ELEVATED TROPONIN I (ACCESS ACCUTNI+3) RESULTS, FOR FOUR PATIENTS, INVOLVING THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. THIS REPORT IS TWO OF FOUR REFERENCING THE PATIENT ON THE EVENT DATE NOTED. THE ELEVATED RESULT WAS RELEASED OUT OF THE LABORATORY. THE CUSTOMER WAS NOT AWARE OF ANY IMPACT TO PATIENT CARE. SUBSEQUENT TESTING OF THE PATIENT'S SAMPLE, ON THE ORIGINAL AND ALTERNATE INSTRUMENT, PRODUCED LOWER RESULTS WITHIN THE NORMAL REFERENCE RANGE. AN AMENDED REPORT WAS ISSUED TO THE HOSPITAL. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE PATIENTS' SAMPLES WERE COLLECTED IN VACUTAINER LITHIUM HEPARIN TUBES AND CENTRIFUGED AT 6,000 RPM (ROTATIONS PER MINUTE) FOR FIVE MINUTES, AT ROOM TEMPERATURE. NO SAMPLE INTEGRITY ISSUES WERE OBSERVED. SYSTEM CHECK, PERFORMED ON (B)(4) 2014, INCLUDING ALL PARAMETERS, WAS WITHIN SPECIFICATIONS. THE CUSTOMER NOTED LOW LEVEL PRECISION TEST FAILED PERCENT COEFFICIENT OF VARIATION (%CV). A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300642 | UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |