FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
MDR report key: 3822268
·
Received May 20, 2014
Report
- Report Number
- 1061932-2014-01122
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 23, 2014
- Report Date
- April 23, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2014, THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND IDENTIFIED THE HEMOGLOBIN CUVETTE REQUIRED REPLACEMENT. THE FSE REPLACED THE HEMOGLOBIN CUVETTE, AND VERIFIED INSTRUMENT PERFORMANCE. (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED AN ERROR CONDITION OF HEMOGLOBIN INCOMPLETE AND FAILURES OF THE DAILY CHECK WHEN USING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300795 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |