FDA Adverse Event Malfunction Summary report: N

UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

MDR report key: 3822268 · Received May 20, 2014

Report

Report Number
1061932-2014-01122
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K081930
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND IDENTIFIED THE HEMOGLOBIN CUVETTE REQUIRED REPLACEMENT. THE FSE REPLACED THE HEMOGLOBIN CUVETTE, AND VERIFIED INSTRUMENT PERFORMANCE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN ERROR CONDITION OF HEMOGLOBIN INCOMPLETE AND FAILURES OF THE DAILY CHECK WHEN USING THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300795 UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1