FDA Adverse Event Malfunction Summary report: N

COULTER LH 500 HEMATOLOGY ANALYZER

MDR report key: 3822266 · Received May 20, 2014

Report

Report Number
1061932-2014-01121
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 24, 2014
Report Date
April 24, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER EVALUATED THE INSTRUMENT AND FOUND THE TUBING AT PV49 WAS CUT AND WAS LEAKING. THE FSE REPLACED THE TUBING AT PV49 AND THE INSTRUMENT RAN WITHOUT ANY LEAKS OR DILUENT COMPARISON ERRORS. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED A LEAK WHEN USING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE INSTRUMENT ALSO GENERATED DILUENT COMPARISON ERRORS. THE VOLUME OF THE LEAK WAS LESS THAN 50CC AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF GLOVES, GLASSES AND A LAB COAT WHEN THE LEAK WAS IDENTIFIED. THERE WAS NO EXPOSURE TO MUCOUS MEMBRANES OR CUTS. THERE WAS NO BIOHAZARD EXPOSURE REPORTED. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300630 COULTER LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1