FDA Adverse Event Malfunction Summary report: N

COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER

MDR report key: 3822264 · Received May 20, 2014

Report

Report Number
1061932-2014-01114
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND FOUND A LEAK AT THE NEEDLE ASSEMBLY CAUSED BY THE NEEDLE BELLOWS OVERFLOWING. THE OVERFLOW WAS CAUSED BY PINCH VALVE (PV21) NOT ACTUATING. THE FSE REPLACED SOLENOID (LV29) AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK IN THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE INSTRUMENT WAS IN SHUTDOWN WHEN THE LEAK WAS NOTICED. THE VOLUME OF THE LEAK WAS ABOUT 2 OZ. AND IT WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF LAB COAT, GLOVES, AND GOGGLES WHEN THE INSTRUMENT WAS CLEANED. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THERE WERE NO ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO THE USER OR PATIENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300629 COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1