PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-84844
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 15, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. CRACKS TO THE CASE, BATTERY TUBE THREADS, RESERVOIR TUBE LIP AND DAMAGES TO THE LCD WINDOW NOTED.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE CUSTOMER WAS EXPERIENCING HIGH BLOOD GLUCOSE OF 19.6MMOL/L. THE CALLER STATED THAT INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS AND THE DEVICE ALARMED AS WELL WAS STUCK IN THE PRIME LOOP. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CALLER STATED THAT ACCIDENTALLY THE DEVICE WAS DROPPED. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300588 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |