FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3822227
·
Received May 20, 2014
Report
- Report Number
- 3004209178-2014-84838
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 17, 2014
- Report Date
- May 8, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST FOLLOWED BY A MOTOR ERROR ALARM DURING THE REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS, AND RESERVOIR TUBE LIP AS WELL SCRATCHED SCREEN.
Description of Event or Problem · 1
THE CUSTOMER REPORTED HAVING AN MRI AND THE INSULIN PUMP WAS NEAR THE DOOR WAY OF THE ROOM. THE BLOOD GLUCOSE READING WAS 250MG/DL. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300680 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR |