ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-03387
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 14, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED WITH THE BLACK TISSUE PAD DAMAGED, MELTED, AND A SMALL PORTION WAS NOT RETURNED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT ACTIVATED DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, A SMALL BLACK PAD FELL OFF FROM THE JAW. THE BLACK PAD WAS RETRIEVED BY FORCEPS. CHANGED TO A NEW DEVICE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. PLEASE SEE THE ENCLOSED PICTURE. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 299450 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |