FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3822005 · Received May 20, 2014

Report

Report Number
1416980-2014-16297
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 25, 2014
Report Date
April 25, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. INTERNAL AND EXTERNAL INSPECTIONS WERE PERFORMED AND THE DEVICE PASSED. TEMPERATURE CONFIRMATION TESTING WAS PERFORMED AND THE DEVICE PASSED. RITE (RETURNED INSTRUMENT TEST EVALUATION) ELECTRICAL TESTING WAS CONDUCTED AND THE DEVICE PASSED. THE DEVICE FAILED RITE FUNCTIONAL TESTING AS FLUID WAS TRANSFERRED ABOVE SPECIFICATION LIMITS. TEST PISTON FOAM WAS INSTALLED, AND THE DEVICE PASSED THE ACCURACY CONFIRMATION TESTING. FURTHER INSPECTION OF THE DOOR ASSEMBLY REVEALED THAT THE DOOR PISTON WAS CRACKED. THE CAUSE OF THE RITE FAILURE WAS DUE TO DETERIORATED PISTON FOAM AND A CRACKED DOOR PISTON. THE PISTON FOAM AND DOOR PISTON WERE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, BUT THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300431 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1