FDA Adverse Event Injury Summary report: N

M2A-MAGNUM MOD HD SZ 54MM

MDR report key: 3821945 · Received May 20, 2014

Report

Report Number
0001825034-2014-04371
Event Type
Injury
Date Received
May 20, 2014
Date of Event
January 9, 2007
Report Date
October 31, 2014
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." NUMBER 4 STATES, ¿LOOSENING OR MIGRATION OF THE IMPLANTS MAY OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, EXCESSIVE ACTIVITY.¿ NUMBER 6 STATES, "INADEQUATE RANGE OF MOTION DUE TO IMPROPER SELECTION OR POSITIONING OF COMPONENTS." NUMBER 9 STATES, "FATIGUE FRACTURE OF COMPONENT MAY OCCUR AS A RESULT OF LOSS OF FIXATION, STRENUOUS ACTIVITY, MALALIGNMENT, TRAUMA, NON-UNION, OR EXCESSIVE WEIGHT." NUMBER 14 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." NUMBER 15 STATES, "ELEVATED METAL ION LEVELS HAVE BEEN REPORTED WITH METAL ON METAL ARTICULATING SURFACES." THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2014-04371 / 04372).

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY CORRECTED AND ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. CORRECTION: INITIAL INFORMATION RECEIVED REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. ADDITIONAL INFORMATION RECEIVED IN PATIENT MEDICAL RECORDS INDICATES THIS IS A REVISION DATE. THE INITIAL IMPLANT DATE IS UNKNOWN.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO CORRECT INFORMATION THAT WAS REPORTED IN ERROR ON THE INITIAL MEDWATCH.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2006. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2007 DUE TO PATIENT ALLEGATIONS OF PAIN, LOSS OF RANGE OF MOTION, NERVE DAMAGE, FOOT DROP, A FRACTURED FEMUR, LOOSENING, ELEVATED METAL ION LEVELS AND METALLOSIS. THE MODULAR HEAD, ACETABULAR CUP AND TAPER ADAPTER WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

LEGAL COUNSEL FOR PATIENT REPORTED PATIENT UNDERWENT A LEFT HIP REVISION ON (B)(6) 2006 OF UNKNOWN PRODUCT. SUBSEQUENTLY, PATIENT'S LEGAL COUNSEL REPORTED PATIENT WAS REVISED ON (B)(6) 2007 DUE TO PATIENT ALLEGATIONS OF PAIN, LOSS OF RANGE OF MOTION, NERVE DAMAGE, FOOT DROP, A FRACTURED FEMUR, LOOSENING, ELEVATED METAL ION LEVELS AND METALLOSIS. THE MODULAR HEAD, ACETABULAR CUP AND TAPER ADAPTER WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. ADDITIONAL INFORMATION RECEIVED IN PATIENT OPERATIVE NOTES FOR THE LEFT REVISION DATED (B)(6) 2007 REPORTS THE PRESENCE OF SCAR TISSUE AND A LOOSE CUP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300145 M2A-MAGNUM MOD HD SZ 54MM PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 545380

Patients

Seq Age Sex Outcome Treatment
1 36 YR Hospitalization| R