ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-03383
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION WITH THE BLADE TIP INTACT AND NOT BROKEN OFF AS REPORTED BY THE CUSTOMER.THE DEVICE WAS TESTED WITH GENERATOR AND WAS FUNCTIONAL, THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. IT IS POSSIBLE THAT THE DEVICE RETURNED MAY HAVE BEEN USED TO COMPLETE THE PROCEDURE AND IS NOT THE DEVICE COMPLAINED ABOUT.
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC LEFT HEPATECTOMY PROCEDURE, THE TITANIUM TIP BROKE. THE BROKEN PIECE WAS RETRIEVED BY FORCEPS. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300131 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |