FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3821939 · Received May 20, 2014

Report

Report Number
3005075853-2014-03383
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION WITH THE BLADE TIP INTACT AND NOT BROKEN OFF AS REPORTED BY THE CUSTOMER.THE DEVICE WAS TESTED WITH GENERATOR AND WAS FUNCTIONAL, THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. IT IS POSSIBLE THAT THE DEVICE RETURNED MAY HAVE BEEN USED TO COMPLETE THE PROCEDURE AND IS NOT THE DEVICE COMPLAINED ABOUT.

Additional Manufacturer Narrative · 1

(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC LEFT HEPATECTOMY PROCEDURE, THE TITANIUM TIP BROKE. THE BROKEN PIECE WAS RETRIEVED BY FORCEPS. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300131 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE