FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3821936 · Received May 20, 2014

Report

Report Number
3005075853-2014-03380
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
May 9, 2014
Report Date
May 13, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED ¿ EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE ¿INSTRUMENT ERROR¿ ALERT WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE TO STOP ACTIVATING AND DISPLAY AN INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿ OR ¿BLADE ERROR DETECTED¿ FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLON- RECTAL CANCER PROCEDURE, DURING THE DISSECTION OF THE PARIETO-COLIC AREA, THE ACTIVE BLADE BROKE OFF. THE PROCEDURE WAS PROLONGED FOR AN UNKNOWN TIME TO RETRIEVE THE LOST BLADE. THE PROCEDURE WAS FINISHED BY USING A NEW DEVICE. NO PATIENT ADVERSE CONSEQUENCES HAVE BEEN REPORTED. PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300130 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE