FDA Adverse Event Injury Summary report: N

EDWARDS RETROFLEX 3/ SAPIEN INTRODUCER SHEATH SET

MDR report key: 3821827 · Received May 20, 2014

Report

Report Number
2015691-2014-01192
Event Type
Injury
Date Received
May 20, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
EDWARDS LIFESCIENCES
Product Code
DYB
PMA / PMN Number
K093877
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), CARDIOVASCULAR COMPLICATIONS, INCLUDING PERFORATION OR DISSECTION OF VESSELS WHICH MAY REQUIRE INTERVENTION, ARE POTENTIAL ADVERSE EVENTS ASSOCIATED WITH THE TRANSFEMORAL TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE. ACCORDING TO LITERATURE REVIEW, AND AS DOCUMENTED IN A TECHNICAL SUMMARY WRITTEN BY EDWARDS LIFESCIENCES, VASCULAR COMPLICATIONS ARE A WELL RECOGNIZED COMPLICATION OF THE TRANSFEMORAL TAVR PROCEDURE IN THIS ELDERLY POPULATION WITH MULTIPLE CO-MORBIDITIES. EDWARDS HAS REVIEWED MANY REPORTS, INCLUDING SCREENING DATA RECORDS AND SOURCE DOCUMENTATION OF VASCULAR COMPLICATIONS AND HAS FOUND THAT THE ROOT CAUSE IS TYPICALLY RELATED TO A COMBINATION OF VESSEL SIZE, TORTUOSITY AND CALCIFICATIONS. ALTHOUGH THE INCIDENCE IS DECREASING WITH SMALLER SHEATH/DELIVERY SYSTEM SIZES AND PHYSICIAN EXPERIENCE, THERE WILL CONTINUE TO BE CASES IN WHICH VASCULAR COMPLICATIONS WILL OCCUR. THE THV PHYSICIAN TRAINING MANUALS INSTRUCT ON PROCEDURAL CONSIDERATIONS FOR SHEATH INSERTION WITH REGARDS TO PROPER SCREENING CRITICAL TO REDUCING VASCULAR COMPLICATIONS. THE TRAINING MANUAL INSTRUCTS THE OPERATOR ON PROPER SHEATH INSERTION AND WITHDRAWAL TECHNIQUES, INCLUDING PRE-DILATING THE VESSEL WITH THE EDWARDS DILATORS. IT ALSO NOTES THAT CALCIFICATION MAY REDUCE LUMEN DIAMETER AND LIMIT OR PREVENT TRANSFEMORAL PASSAGE OF THE DEVICES. THE IFU CONTRAINDICATES PATIENTS WITH ILIO-FEMORAL VESSEL CHARACTERISTICS THAT WOULD PRECLUDE SAFE PLACEMENT OF SHEATHS SUCH AS SEVERE OBSTRUCTIVE CALCIFICATION OR SEVERE TORTUOSITY. PRE-PROCEDURE SCREENING AND ASSESSMENT OF THE FEMORAL/ILIAC ARTERY INTERNAL DIAMETERS WILL ENABLE THE CLINICIAN TO DETERMINE IF THE SAPIEN VALVE CAN BE DELIVERED TRANSFEMORALLY. ASSESSMENT OF LOCATION AND AMOUNT OF CIRCUMFERENTIAL CALCIUM WILL AID IN DETERMINING AREAS OF REDUCED VESSEL DIAMETERS. THE OPERATORS ARE TRAINED TO MEASURE MINIMUM VESSEL DIAMETER TAKING CALCIUM INTO ACCOUNT. THE PHYSICIAN TRAINING MANUAL ALSO LISTS THE MINIMUM RECOMMENDED VESSEL SIZE FOR EACH SIZE DEVICE. DESPITE THE BEST SCREENING TOOLS, A SMALL PERCENTAGE OF PATIENTS WILL HAVE FEMORAL/ILIAC VESSELS THAT ARE NOT AMENABLE TO THE TRANS-FEMORAL APPROACH OR WHERE INCREASED RESISTANCE IS ENCOUNTERED DURING INSERTION OF DEVICES. IN MANY CASES, THE VESSEL MINIMUM LUMINAL DIAMETER (MLD) MAY BE BORDERLINE OR BELOW THE INDICATED SIZE. IN ADDITION, SIGNIFICANT CALCIFICATION AND/OR TORTUOSITY, NOT ALWAYS APPRECIABLE ON IMAGING, COULD BE CONTRIBUTING FACTORS TO THE EVENT. THE MINIMUM REQUIRED VESSEL DIAMETER FOR A 24FR RF3 SHEATH IS >8MM. IN THIS CASE, THE PATIENT¿S ACCESS VESSEL MLD WAS 8MM WITH MILD TORTUOSITY, MODERATE CALCIFICATION IN THE LCFA AND SEVERE CALCIFICATION IN THE CIA NOTED. IT APPEARS THAT PATIENT FACTORS MAY HAVE CONTRIBUTED TO THIS EVENT. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A (B)(4) BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS (B)(4) REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.

Description of Event or Problem · 1

DURING THE TF TAVR, MODERATE DIFFICULTY WAS ENCOUNTERED WHILE INSERTING THE 24FR RETROFLEX 3 SHEATH. FRICTION WAS FELT WHEN THE SHEATH PASSED THROUGH THE COMMON ILIAC ARTERY. INSERTION DIFFICULTY WAS ALSO ENCOUNTERED DURING THE INSERTION OF THE DELIVERY SYSTEM AND VALVE THROUGH THE SHEATH. THE PHYSICIAN PUSHED ¿HARD¿ AND THE DELIVERY SYSTEM POPPED THOUGH WHAT APPEARED TO BE A KINK IN THE SHEATH. FOLLOWING THE SUCCESSFUL DEPLOYMENT OF THE VALVE, THE SHEATH WAS REMOVED AND THE ACCESS SITE WAS REPAIRED. FOLLOWING THE REMOVAL OF THE SHEATH, A PLAQUE DISRUPTION AT THE INSERTION SITE WAS NOTED. AN ENDARTERECTOMY AND PLEDGETED SUTURE REPAIR WERE PERFORMED. DOPPLER REVEALED IMPROVEMENT OF FLOW TO THE EXTREMITY. THE PATIENT WAS TRANSFERRED TO RECOVERY IN STABLE CONDITION AND AT LAST REPORT, DOING WELL. THE PATIENT¿S VESSEL MINIMUM LUMINAL DIAMETER (MLD) WAS 8MM, VESSEL TORTUOSITY WAS NOTED TO BE MILD. THE PATIENT¿S LEFT COMMON FEMORAL ARTERY (LCFA) WAS REPORTED TO BE MODERATELY CALCIFIED AND THE COMMON ILIAC ARTERY (CIA) WAS REPORTED TO BE HEAVILY CALCIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
300293 EDWARDS RETROFLEX 3/ SAPIEN INTRODUCER SHEATH SET INTRODUCER, CATHETER DYB EDWARDS LIFESCIENCES 9120S26 59536617

Patients

Seq Age Sex Outcome Treatment
1 91 YR Required Intervention