FENESTRATED BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-03104
- Event Type
- Malfunction
- Date Received
- May 20, 2014
- Date of Event
- April 23, 2014
- Report Date
- April 24, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A FRAYED PITCH CABLE AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT IN CENTRAL PROCESSING, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS OBSERVED TO HAVE 'WIRES STARTING TO FRAY. THE INSTRUMENT WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT. NOTHING REPORTEDLY FELL INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 298464 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | N10140106 662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |