FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3821529 · Received May 20, 2014

Report

Report Number
2024168-2014-03192
Event Type
Injury
Date Received
May 20, 2014
Date of Event
May 2, 2014
Report Date
May 7, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THE LOT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE SAFETY AND EFFECTIVENESS OF THE PERCLOSE PROGLIDE DEVICES HAVE NOT BEEN ESTABLISHED IN THE FOLLOWING PATIENT POPULATIONS: PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. THE DEVICE WAS REPORTED TO BE DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER PROGLIDE DEVICE REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE HEAVILY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 5FR SHEATH, AFTER AN INTERVENTIONAL PROCEDURE FOR PERIPHERAL ARTERIAL OCCLUSIVE DISEASE (PAOD). REPORTEDLY, AN ANTERIOR AND POSTERIOR NEEDLE TO CUFF MISS OCCURRED AFTER THE PLUNGER WAS REMOVED. A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299822 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 30930K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 5F, HEPARIN