FDA Adverse Event Malfunction Summary report: N

MONOMAX VIOLET 1 (4) 150CM HRT48 LOOP (M)

MDR report key: 3821473 · Received January 15, 2014

Report

Report Number
2916714-2014-00034
Event Type
Malfunction
Date Received
January 15, 2014
Report Date
January 15, 2014
Manufacturer
B.BRAUN SURGICAL S.A.
Product Code
NWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

US REPORTING AGENT NOTIFIED ON (B)(4) 2013. REPORTED DEVICE NOT MARKETED IN THE US; HOWEVER, SIMILAR DEVICE IS. MFG SITE EVAL: SAMPLES RECEIVED: 1 UNOPENED BOX WITH 24 POUCHES. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL OEM REQUIREMENTS. THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH. THERE ARE NO UNITS IN OEM STOCK. ALL PACKS RECEIVED ARE TIGHT. KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED WAS TESTED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE OEM. KNOT PULL TENSILE STRENGTH RESULTS BEFORE RELEASING THE PRODUCT WERE IN ACCORDANCE WITH OEM SPECIFICATIONS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED).

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). THREAD BREAKS OFTEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40431 MONOMAX VIOLET 1 (4) 150CM HRT48 LOOP (M) NONE NWJ B.BRAUN SURGICAL S.A. B0041222 112163V004

Patients

Seq Age Sex Outcome Treatment
1