FDA Adverse Event Malfunction Summary report: N

CATGUT PLAIN 5 (9) 25M

MDR report key: 3821467 · Received January 14, 2014

Report

Report Number
2916714-2014-00027
Event Type
Malfunction
Date Received
January 14, 2014
Report Date
January 14, 2014
Manufacturer
B. BRAUN SURGICAL S.A.
Product Code
GAL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

US REPORTING AGENT NOTIFIED ON 12/20/2013. REPORTED DEVICE NOT MARKETED IN THE US; HOWEVER, SIMILAR DEVICE IS. MANUFACTURING SITE EVALUATION: SAMPLES RECEIVED: 1 CASSETTE. REVIEWED THE BATCH MANUFACTURING RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFIL OEM REQUIREMENTS. THERE ARE NO UNITS OF THIS BATCH NUMBER IN OEM STOCK. KNOT PULL TENSILE STRENGTH OF THE SAMPLES RECEIVED WAS TESTED AND THE RESULTS FULFILL THE REQUIREMENTS OF THE OEM. ABSORPTION CONTROLS CONDUCTED ON SAMPLE WITH COLLAGENASE AFTER A PERIOD OF TIME HAVE RETENTION STRENGTH VALUES HIGHER THAN CURRENT ONES FOR THIS MATERIAL AND SIZE. THEREFORE, THIS THREAD DOES NOT DEGRADE QUICKER THAN OTHER BATCHES TESTED. THE THREAD OF THE SAMPLE RECEIVED WAS ALMOST CONSUMED AND WE COULD NOT CONDUCT ALL DEGRADATION TESTS WITH COLLAGENASE IN ORDER TO VIEW THE DEGRADATION PROFILE. THEREFORE, AND ACCORDING TO THE PERFORMED ANALYSIS RESULTS, OEM CONSIDERS THIS COMPLAINT AS NOT CORRESPONDING. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED). ACTIONS ON PRODUCT: NOT APPLICABLE. CORRECTIVE/PREVENTIVE ACTIONS: NOT APPLICABLE.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). TYPE OF OPERATION: SECTION/CAESARIAN. TISSUE BEING SUTURED: UTERUS. SUTURE TECHNIQUE: CONTINUOUS. DESCRIPTION: (B)(6) 2013 SECTION CAESAREAN, (B)(6) 2013 COW STOPPED TO EAT, SURGICAL INTERVENTION, RUPTURE OF THE CATGUT IN UTERUS SUTURE. FINALLY THEY EUTHANIZED THE COW ON (B)(6) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
37508 CATGUT PLAIN 5 (9) 25M NONE GAL B. BRAUN SURGICAL S.A. F1144111 611363

Patients

Seq Age Sex Outcome Treatment
1 Other