FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR(R) PLUS DEFIBRILLATOR

MDR report key: 3821444 · Received May 20, 2014

Report

Report Number
3015876-2014-00569
Event Type
Malfunction
Date Received
May 20, 2014
Date of Event
April 29, 2014
Report Date
April 29, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K033275
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE CUSTOMER REGARDING THE STATUS OF THEIR DEVICE BUT HAS NOT BEEN SUCCESSFUL. IT IS UNKNOWN IF THE CUSTOMER WILL BE REPLACING THE DEVICE OR RETURNING IT TO PHYSIO-CONTROL FOR EVALUATION. THE DEVICE HAS NOT BEEN RETURNED TO PHYSIO-CONTROL. THE CAUSE OF THE REPORTED FAILURE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD ALL THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) PRESENT ON THE DISPLAY AND THE UNIT WILL NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299965 LIFEPAK CR(R) PLUS DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC CRPLUS

Patients

Seq Age Sex Outcome Treatment
1