FDA Adverse Event Malfunction Summary report: N

CENTURY BED

MDR report key: 3821374 · Received April 28, 2014

Report

Report Number
1824206-2014-01321
Event Type
Malfunction
Date Received
April 28, 2014
Date of Event
March 31, 2014
Report Date
March 31, 2014
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM TECHNICAL SUPPORT FOUND THE BRAKE CASTERS NEEDED TO BE ADJUSTED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. HILL-ROM TECHNICAL SUPPORT RECOMMENDED ADJUSTING THE BRAKE CASTER. THREE ATTEMPTS HAVE BEEN MADE REGARDING A RESOLUTION TO THIS CONTACT LINE, WITH NO RESPONSE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN THE PATIENT ROOM AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254451 CENTURY BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 835

Patients

Seq Age Sex Outcome Treatment
1