VERSACARE BED
Report
- Report Number
- 1824206-2014-01328
- Event Type
- Malfunction
- Date Received
- April 28, 2014
- Date of Event
- March 31, 2014
- Report Date
- March 31, 2014
- Manufacturer
- HILL-ROM INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE TECHNICIAN FOUND THE SIDE RAIL CENTER ARM ASSEMBLY WAS BROKEN. PER THE HILL-ROM SERVICE MANUAL, PERFORM ANNUAL PREVENTIVE MAINTENANCE PROCEDURES TO MAKE SURE ALL VERSACARE BED COMPONENTS ARE FUNCTIONING AS ORIGINALLY DESIGNED. MAKE SURE THE HEAD AND INTERMEDIATE SIDE RAILS ARE NOT BENT OR TWISTED. MAKE SURE THE SIDE RAILS LOCK CORRECTLY IN THE HIGH POSITION AND YOU HEAR THE LATCH CLICK. GENTLY PULL ON EACH SIDE RAIL TO MAKE SURE IT LATCH COMPONENTS ARE CLEAN AND LOOK FOR OBSTRUCTIONS. RELEASE EACH SIDE RAIL FROM THE UP POSITION, AND LET IT FALL FREELY. THE SIDE RAIL SHOULD LOWER SLOWLY AND SMOOTHLY. REMOVE THE CENTER ARM COVER, AND EXAMINE THE CABLE ROUTING FOR PINCHING, BINDING, OR DAMAGE. MAKE SURE THE LATCH MECHANISM OF EACH SIDE RAIL IS IN GOOD CONDITION. REMOVE THE SIDE RAIL COVER, AND MAKE SURE THE MOUNTING SCREWS ARE FULLY INSTALLED. MAKE SURE ALL FUNCTIONS ON THE CAREGIVER CONTROL WORK CORRECTLY. REPAIR OR REPLACE THE SIE RAILS AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2010 AND 2012 TO 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SIDE RAIL CENTER ARM ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE SIDE RAIL WOULD NOT LATCH. THE BED WAS LOCATED IN THE BASEMENT OF THE ACCOUNT. THERE WAS NO PATIENT/USER REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254532 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |