FDA Adverse Event Malfunction Summary report: N

ADVNATA BED

MDR report key: 3821259 · Received April 28, 2014

Report

Report Number
1824206-2014-01318
Event Type
Malfunction
Date Received
April 28, 2014
Date of Event
March 31, 2014
Report Date
March 31, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BRAKES NOT HOLDING. PER THE HILL-ROM SERVICE MANUAL THE ADVANTA BED REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE BRAKE CASTERS TO DETERMINE IF THE BED MOVES WHEN YOU ACTIVATE THE BRAKE MODE. TEST THE STEER MODE TO DETERMINE IF THE FOOT AND CASTERS LOCK IN THE STEER MODE. CHECK THE TIRES FOR CUTS, WEAR, TREAD LIFE, ETC. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED FROM 2006 TO 2011 AND 2013 TO 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH REPLACED THE BRAKE LINKAGE, SCREWS AND CROSS SHAFT TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NO HOLDING. THE BED WAS LOCATED IN THE BED SHOP AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254458 ADVNATA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1