FDA Adverse Event Malfunction Summary report: N

CARE ASSIST BED

MDR report key: 3821249 · Received April 28, 2014

Report

Report Number
9615739-2014-00175
Event Type
Malfunction
Date Received
April 28, 2014
Date of Event
March 31, 2014
Report Date
March 31, 2014
Manufacturer
HILL-ROM PLUVINGER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE BRAKE BAR SPACER, CASTER SUPPORT AND BRAKE SWITCH WERE BROKEN. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR, AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE PEDALS ARE PRESSED AND REPAIR IS NECESSARY. A SEARCH OF THE HILL-ROM RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 AND 2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH REPLACED THE BRAKE BAR SPACER, CASTER SUPPORT NAD BRAKE SWITCH TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NO HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN THE HCA WAREHOUSE. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254423 CARE ASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVINGER 1170

Patients

Seq Age Sex Outcome Treatment
1