FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3821235 · Received April 16, 2014

Report

Report Number
1720753-2014-03305
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
March 31, 2014
Report Date
April 16, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE BACKUP BATTERY ON GPOS WAS EVALUATED AND REPLACED. THE CONNECTION OF RUI CONSOLE WAS EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FROZE UP AND THEN WOULD NOT PERFORM FLUOROSCOPY. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233635 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1