LIFEVEST WCD 4000 SYSTEM

MDR report key: 3821232 · Received April 29, 2014

Report

Report Number
3008642652-2014-01244
Event Type
Malfunction
Date Received
April 29, 2014
Date of Event
March 26, 2014
Report Date
April 21, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEMS (CHECK THERAPY PAD MESSAGES AND DAMAGED CABLE OR WIRES EXPOSED) WERE CONFIRMED. UPON INVESTIGATION THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND DN. THE ROOT CAUSE OF THE PULLED CABLES COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CHECK THERAPY PAD MESSAGES AND A DAMAGE CABLE. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.