FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3821181 · Received April 16, 2014

Report

Report Number
1720753-2014-03320
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
April 3, 2014
Report Date
April 16, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE X-RAY TUBE WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HIGH VOLTAGE ARCING FROM THE SYSTEM FOLLOWED BY A LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233551 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1