FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNE
MDR report key: 3821135
·
Received April 16, 2014
Report
- Report Number
- 9615050-2014-02741
- Event Type
- Malfunction
- Date Received
- April 16, 2014
- Date of Event
- January 1, 2014
- Report Date
- March 17, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0069-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THE TOUCH SCREEN WILL NOT CALIBRATE. THE DEVICE WAS IN THE BIOMEDICAL DEPT WITH A NOTE THAT INDICATED THE TOUCHSCREEN WILL NOT CALIBRATE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232536 | SYMBIQ SINGLE CHANNE | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |