FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNE

MDR report key: 3821135 · Received April 16, 2014

Report

Report Number
9615050-2014-02741
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
January 1, 2014
Report Date
March 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE TOUCH SCREEN WILL NOT CALIBRATE. THE DEVICE WAS IN THE BIOMEDICAL DEPT WITH A NOTE THAT INDICATED THE TOUCHSCREEN WILL NOT CALIBRATE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO BE OUT OF CALIBRATION. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232536 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA