FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNE

MDR report key: 3821126 · Received April 16, 2014

Report

Report Number
9615050-2014-02730
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
March 1, 2014
Report Date
March 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION AT THE SERVICE CENTER FOUND THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THE PROBABLE CAUSE WAS CORROSION ON THE BOTTOM EDGE OF THE TOUCHSCREEN AND FRONT "BEZED" DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED DURING PROGRAMMING PRIOR TO PT USE THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THE CUSTOMER CONTACT REPORTED DURING TESTING AT THE USER FACILITY THE TOUCHSCREEN DID NOT RESPOND AND NO FURTHER TESTING COULD BE COMPLETED. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232533 SYMBIQ SINGLE CHANNE 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA