SYMBIQ SINGLE CHANNE
Report
- Report Number
- 9615050-2014-02730
- Event Type
- Malfunction
- Date Received
- April 16, 2014
- Date of Event
- March 1, 2014
- Report Date
- March 17, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0069-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION AT THE SERVICE CENTER FOUND THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THE PROBABLE CAUSE WAS CORROSION ON THE BOTTOM EDGE OF THE TOUCHSCREEN AND FRONT "BEZED" DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED DURING PROGRAMMING PRIOR TO PT USE THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THE CUSTOMER CONTACT REPORTED DURING TESTING AT THE USER FACILITY THE TOUCHSCREEN DID NOT RESPOND AND NO FURTHER TESTING COULD BE COMPLETED. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232533 | SYMBIQ SINGLE CHANNE | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |