FDA Adverse Event Malfunction Summary report: N

SYMBIQ SCHANNEL 3.01

MDR report key: 3821125 · Received April 16, 2014

Report

Report Number
9615050-2014-02729
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
March 1, 2014
Report Date
March 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE TOUCHSCREEN WAS FOUND TO NOT RESPOND WHEN PRESSED AND WOULD NOT CALIBRATE. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN WOULD NOT CALIBRATE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT STATED, "TOUCHSCREEN WILL NOT CALIBRATE." NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED AND WOULD NOT CALIBRATE. NO ADD'L INFO WAS PROVIDED.