FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3821120 · Received April 16, 2014

Report

Report Number
1720753-2014-03325
Event Type
Malfunction
Date Received
April 16, 2014
Date of Event
March 28, 2014
Report Date
April 16, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE MAINFRAME X-RAY SWITCH WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT X-RAYS STAY ON AFTER RELEASING THE X-RAY SWITCH. THIS EVENT OCCURRED DURING A PT CARE CASE. NO PT INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232685 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1