FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3821099
·
Received April 16, 2014
Report
- Report Number
- 1720753-2014-03333
- Event Type
- Malfunction
- Date Received
- April 16, 2014
- Date of Event
- March 31, 2014
- Report Date
- April 16, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MAINFRAME GENERATOR BATTERIES WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232630 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |