FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3820840 · Received May 20, 2014

Report

Report Number
2531779-2014-14206
Event Type
Injury
Date Received
May 20, 2014
Date of Event
May 16, 2014
Report Date
May 17, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 07/09/2014. DEVICE EVALUATION: THE DEVICE HAS NOT BEEN RETURNED; A RETAINED SAMPLE FROM THE SAME CARTRIDGE LOT WAS PULLED AND EVALUATED BY PRODUCT ANALYSIS ON 06/30/2014 WITH THE FOLLOWING FINDINGS:A CARTRIDGE FILL TEST WAS COMPLETED WITH NO AIR BUBBLES FORMED. A CARTRIDGE LEAK TEST WAS PERFORMED WITH NOTED LEAKS FROM ANYWHERE ON THE CARTRIDGE. A CARTRIDGE FORCE TEST WAS PERFORMED AND CONFIRMED THAT THE REQUIRED FORCES WERE WITHIN SPECIFICATIONS. NO DEFECTS WERE FOUND RELATED TO THE LOSS OF PRIME COMPLAINT.

Description of Event or Problem · 1

ON (B)(6) 2014 THE REPORTER CONTACTED ANIMAS ALLEGING A BLOOD GLUCOSE OF 2.6 MMOL/L WITH DIZZINESS RELATED TO MULTIPLE LOSS OF PRIME WARNINGS ON THE PUMP. THE REPORTER INDICATED THAT THE PUMP EMITTED MANY LOSS OF PRIME WARNINGS REQUIRING REPRIMING OF THE PUMP OCCURRING ON A SINGLE CARTRIDGE. DURING THE REVIEW OF THE PUMP IT WAS FOUND THAT THE PATIENT HAD PRIMED 19.2 UNITS TOTAL BETWEEN (B)(6) 2014 AND THE PATIENT INDICATED THAT THE PUMP WAS NOT DISCONNECTED DURING THE PRIME. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPOGLYCEMIA RELATED TO INCORRECTLY PRIMING WHILE ATTACHED FOR MULTIPLE LOSS OF PRIME WARNINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299775 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR Life Threatening