PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG
Report
- Report Number
- 1419937-2014-00362
- Event Type
- Malfunction
- Date Received
- April 14, 2014
- Date of Event
- April 8, 2014
- Report Date
- April 8, 2014
- Manufacturer
- MEDELA, INC.
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A REPLACEMENT TRANSFORMER WAS SENT TO THE CUSTOMER. THE PRODUCT INVOLVED IN THE COMPLAINT WAS NOT RETURNED FOR EVALUATION/INVESTIGATION AT THIS TIME. THEREFORE, NO CONCLUSIONS CAN BE MADE AS TO THE CAUSE OF THE EVENT. ATTEMPTS TO RETRIEVE THE PRODUCT ARE ONGOING. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FALL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. SHOULD ADDITIONAL INFORMATION OR THE ORIGINAL PRODUCT BE RECEIVED, RESULTING IN NEW, CHANGED, OR CORRECTED INFORMATION, A FOLLOW UP REPORT WILL BE FILED AT THAT TIME.
THE CUSTOMER REPORTED THAT THE HOUSING WAS BROKEN, WHICH IS A SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 228255 | PUMP IN STYLE ADVANCED BREAST PUMP METRO BAG | HGX | MEDELA, INC. | 57036/9207010 | REV L-2011 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |