FDA Adverse Event Malfunction Summary report: N

2.4MM RECONSTRUCTION, 14 HOLE STRAIGHT LOCKING

MDR report key: 3820730 · Received May 1, 2014

Report

Report Number
2027754-2014-00006
Event Type
Malfunction
Date Received
May 1, 2014
Date of Event
April 9, 2014
Report Date
April 28, 2014
Manufacturer
OSTEOMED
Product Code
JEY
PMA / PMN Number
K080694
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RETURNED PRODUCT WAS PIECED TOGETHER TO SHOW THE 34 DEGREES BEND OF THE STRAIGHT PLATE. DOCTOR HAS CONTROL OVER CONTOURING PLATE TO DESIRED SHAPE HOWEVER THE 34 DEGREES ANGLE MAY HAVE CAUSED A WEAK SPOT IN THE BEND AREA. IT MAY HAVE BEEN BETTER TO BEND OVER A THREE HOLE SPAN RATHER THAN A SINGLE HOLE SPAN. IT IS LIKELY THAT THE PLATE MAY HAVE HAD A SMALL CRACK FORM WHEN IT WAS CONTOURED TO FIT PATIENT ANATOMY. THIS CRACK WOULD HAVE PROPAGATED WITH THE CYCLIC LOADING THAT IT SAW AS THE PATIENT STARTED EATING AGAIN DURING RECOVERY. WHY DOCTOR WAITED OVER TWO MONTHS TO EXPLANT IS UNKNOWN. GENERAL SURGICAL TECHNIQUE GUIDES POINT TO VARIOUS TECHNIQUE TIPS AND NOTES TO ASSIST IN OBTAINING THE BEST BENDS.

Description of Event or Problem · 1

PATIENT WAS TREATED FOR A FRACTURE AT THE ANGLE OF THE MANDIBLE WITH OSTEOMED PART NUMBER 214-2417. SOMETIME BETWEEN THE INITIAL SURGERY WHICH TOOK PLACE ON (B)(6) 2013 AND (B)(6)2014 THE RECONSTRUCTION PLATE THAT WAS PLACED BROKE. (B)(6) WAS WHEN THE DOCTOR NOTICED THE FRACTURE. EXPLANT OF THE BROKEN PLATE TOOK PLACE ON (B)(6) 2014. THE CURRENT STATUS OF THE PATIENT IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262416 2.4MM RECONSTRUCTION, 14 HOLE STRAIGHT LOCKING 2.4 MM RECONSTRUCTION PLATE JEY OSTEOMED

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention