2.4MM RECONSTRUCTION, 14 HOLE STRAIGHT LOCKING
Report
- Report Number
- 2027754-2014-00006
- Event Type
- Malfunction
- Date Received
- May 1, 2014
- Date of Event
- April 9, 2014
- Report Date
- April 28, 2014
- Manufacturer
- OSTEOMED
- Product Code
- JEY
- PMA / PMN Number
- K080694
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
RETURNED PRODUCT WAS PIECED TOGETHER TO SHOW THE 34 DEGREES BEND OF THE STRAIGHT PLATE. DOCTOR HAS CONTROL OVER CONTOURING PLATE TO DESIRED SHAPE HOWEVER THE 34 DEGREES ANGLE MAY HAVE CAUSED A WEAK SPOT IN THE BEND AREA. IT MAY HAVE BEEN BETTER TO BEND OVER A THREE HOLE SPAN RATHER THAN A SINGLE HOLE SPAN. IT IS LIKELY THAT THE PLATE MAY HAVE HAD A SMALL CRACK FORM WHEN IT WAS CONTOURED TO FIT PATIENT ANATOMY. THIS CRACK WOULD HAVE PROPAGATED WITH THE CYCLIC LOADING THAT IT SAW AS THE PATIENT STARTED EATING AGAIN DURING RECOVERY. WHY DOCTOR WAITED OVER TWO MONTHS TO EXPLANT IS UNKNOWN. GENERAL SURGICAL TECHNIQUE GUIDES POINT TO VARIOUS TECHNIQUE TIPS AND NOTES TO ASSIST IN OBTAINING THE BEST BENDS.
PATIENT WAS TREATED FOR A FRACTURE AT THE ANGLE OF THE MANDIBLE WITH OSTEOMED PART NUMBER 214-2417. SOMETIME BETWEEN THE INITIAL SURGERY WHICH TOOK PLACE ON (B)(6) 2013 AND (B)(6)2014 THE RECONSTRUCTION PLATE THAT WAS PLACED BROKE. (B)(6) WAS WHEN THE DOCTOR NOTICED THE FRACTURE. EXPLANT OF THE BROKEN PLATE TOOK PLACE ON (B)(6) 2014. THE CURRENT STATUS OF THE PATIENT IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 262416 | 2.4MM RECONSTRUCTION, 14 HOLE STRAIGHT LOCKING | 2.4 MM RECONSTRUCTION PLATE | JEY | OSTEOMED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Required Intervention |