FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 3820499 · Received April 9, 2014

Report

Report Number
1720753-2014-03100
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
March 19, 2014
Report Date
April 9, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FUSES WERE RESEATED AND THE POWER SUPPLY WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE FE REPORTED THE SYSTEM FAILED TO BOOT UP. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214046 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1