FDA Adverse Event Malfunction Summary report: N

GEMSTAR PAIN MGE SNG

MDR report key: 3820490 · Received April 9, 2014

Report

Report Number
9615050-2014-02573
Event Type
Malfunction
Date Received
April 9, 2014
Date of Event
February 4, 2014
Report Date
February 14, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1159-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, AT POWER UP THE DEVICE ALARMED FOR 18/000 (HISTORY POINTER ERROR) AND THEN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF THE DEVICE ALARMING WITH A 18/000/000 ERROR CODE. NO ADDITIONAL INFO WAS PROVIDED. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE ALARMED WITH A 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214043 GEMSTAR PAIN MGE SNG 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA