FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3820406
·
Received April 9, 2014
Report
- Report Number
- 1720753-2014-03084
- Event Type
- Malfunction
- Date Received
- April 9, 2014
- Date of Event
- March 18, 2014
- Report Date
- April 9, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE IMAGE ON THE LEFT MONITOR WOULD INTERMITTENTLY DISAPPEAR. THIS RESULTED IN AN INTERMITTENT RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214251 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |