Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: UNKNOWN. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. (B)(4). THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: SERVICE HISTORY REVIEW OF THE DEVICE HAS BEEN PERFORMED. THE REVIEW INDICATES THAT THE DEVICE HAS NOT BEEN SERVICED DURING THE PAST 6 MONTHS. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINT ISSUE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE INVESTIGATION IS BASED ON THE SERVICE & REPAIR RECORD (SRR) RECEIVED. THE DEVICE FAILURE REPORTED BY THE CUSTOMER COULD BE CONFIRMED. THE SERVICE HISTORY REVIEW SHOWS NO PREVIOUS SERVICE CONDITIONS RELEVANT TO THE CURRENT COMPLAINT ISSUE. THE SERVICE TECHNICIAN NOTED THE FOLLOWING ACTION TAKEN: REPAIR. THE DEVICE HAS BEEN SERVICED AND FUNCTIONAL TESTING HAS BEEN PERFORMED IN ACCORDANCE WITH SERVICE MANUAL 05.001.201 - PRE-REPAIR DIAGNOSTIC ASSESSMENT/REPAIR VERIFICATION.. DURING THE PRE-REPAIR DIAGNOSTIC ASSESSMENT THE SERVICE TECHNICIAN IDENTIFIED THE FOLLOWING FAILURE: TRIGGER / SLIDER BLOCKED, JAMMED. THE TECHNICIAN INCLUDED THE FOLLOWING COMMENTS ON THE SRR: "HANDLE STICKS" AND "SERVICE DUE" THE PROBABLE ROOT CAUSE OF THE EVENT WAS DETERMINED TO BE NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ONE OF THE BUTTONS/SWITCHES OF THE MODULAR DRIVE IS GETTING STUCK WHEN PRESSED. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
299274 BATTERY HANDPIECE/MODULAR FOR TRAUMA RECON SYSTEM MOTOR, SURGICAL INSTRUMENT, AC-POWERED GEY DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1